1. Supports saving and recalling configuration files under remote control.
2. Supports configuration file upgrade under Webserver.
3. Bug fixes
a. The measurement result save time is incorrect under the scan card.
b. The trend chart is faulty under the scan card.
c. Trigger and acquisition errors occur under the scan card.
d. ACI/Freq and ACI/period switching are not supported under dual display.
e. The center value setting is abnormal under the histogram.
f. After *RST, dual display configuration and statistics reset are abnormal.
1. Supports RE model.
2. Solve the problem of inaccurate measurement values under the 200V range of DCV.
3. Bug fixes
a) Restore the default trigger source error under Cont function.
b) Opening the statistics menu causes the probehold menu to disappear.
c) Optimize the prompt window under the Data Log menu and the trend chart display issue.
d) When the Power On mode is Last, the switch between the temp and cont functions is abnormal.
e) The number of sampling points exceeds 2M and the timestamp is incorrect.
f) When the trigger source is an external trigger, the trigger status changes to automatic after the abort instruction.
g) Support the transmission of scanner card data timestamps.
h) Setting the Socket parameter to MAX causes the system to freeze.
i) Remotely enable relative values, send *RST, and switch to local trigger without reading.
j) The sampling interval of the Data Log cannot be set to 3600s.
k) When the DCI range is switched, the range obtained by the instruction is inconsistent with the UI.
Features and Enhancements:
Bug Fixes:
Features and Enhancements:
1. Added harmonic analysis measurement function in SA measurement mode.
2. Added full-channel calibration function.
3. Added calibration information display function.
4. Added power compensation function for fixture simulation.
5. Added ECal (Electronic Calibration) characterization function.
6. Added Compression Point Search / Saturation Point Search / Normal Operating Point Search functions.
7. Added Preferences settings.
8. TDR Measurement Mode: Now supports extended port function.
9. AFR (Automatic Fixture Removal): Now supports timedomain trace display.
10. Minimum Sweep Points: Now supports a minimum setting of 1 point.
11. File Manager: Added support for “Favorites” and “History” functions.
12. Power Meter: Added support for IP connection.
13. Power Meter Reading: Default frequency changed to 50 MHz.
14. Switch Matrix: Added support for characterization and deembedding functions.
Bug Fixes:
1. Fixed the issue where single-ended measurements (e.g., Sss11) would not sweep under balanced measurement mode.
2. Fixed the issue where smoothing results were abnormal under the Phase format.
3. Fixed the issue where traces might appear abnormal after executing a calibration data overwrite.
4. Fixed the issue where the power calibration progress bar might display incorrectly.
5. Fixed the issue where trace data became abnormal after recalling CSV files saved from Power Sweeps or CW Sweeps.
6. Fixed the issue where the sweep would hang when the external trigger signal frequency was high in “Point Trigger” mode.
7. Fixed the issue where the log display time might be incorrect.
8. Fixed abnormal UI control behavior in the SA measurement mode configuration dialog.
9. Fixed the issue where eye diagram data might be abnormal after modifying the rise time in TDR measurement mode.
1.Optimized Multitone function
2.AWG Added single waveform setting UI
3.Resolved an issue where the UI would occasionally fail to display during startup.
4.Resolved an issue where startup would occasionally fail.
5.Fixed an output error when manually triggering in IQ mode.
Features and Enhancements:
Bug Fixes:
Features and Enhancements:
Bug Fixes:
1. Optimized UI response in the case of long waveform depth
2. Added Spectrum Analysis mode
3.Channel: in Eres mode to display the corresponding bandwidth information in the descriptor box
4.Decode: Added SPMI, Spacewire, 8B/10B, MIPI-RFFE
5.Measure:
a)Added item: Bit Rate, Data rate, △TimeMax, △TimeMin
b)Supported to limit the display digits of measure
6.Memory: Supported to import a file with data of multi-channel
7.Analysis:
a)Eye /Jitter upgraded to 2.0: optimized UI; added explicit clock; supported Equalization; supported Noise analysis; supported differential source; supported F1-F4, M1-M4 as source
b)Compliance Test: added 10BASE-T, 10BASE-T1S, DDR4, PCIe 1.0/2.0
c)Serial Signal Test: added I2C, SPI
d)Bode plot: added crossover frequency measure
8. Probe: Supported SAP8000D probe
9.Roll:
a)Supported Math and track cursors in Stop mode
b)Supported bandwidth limit
10.Webserver: supported inverted screenshot (according to the Image style setting in the Save menu)
11.Save/recall: File Manager: Supported to Select All; Supported to save all frames to a single file in Sequence mode
12.UI:Supported to specify Zoom region by drag a rectangle
13.Utility:Added English help; Supported password for screen saver
14. Fixed several bugs
a) When Log horizontal is used in FFT numbers are scrambled
b) Vertical axis of FFT optimized
c) Grid not aligned with numbers/wrong place with large offset
d)100BASE-T1 compliance test: displayed units of test result and judgment criteria are inconsistent
e) 100Base-TX and 1000Base-T: algorithm optimized
f) USB 2.0: compatible issue with USB IF recommended fixture
g) 1000BASE-T: transmission distortion test automatically executes without confirmation after completing one pair under the pairALL differential probe test
Compatible with hardware version with 32 GB internal storage
1.Optimize vector signal quality
2.HW Information displays the hardware serial number.
3.Remove “None” from the Custom filter type.
4.Added filter settings to ARB waveform settings.
5.Remove filter settings in ARB settings.
6.Added armed auto mode to ARB trigger.
7.The maximum number of multitone has been increased to 25001.
8.Supports playback of GSM, DAB, and WCDMA protocol waveforms.
9.Added automatic startup for SMB server and supports setting SMB server password.
10.Fixed several bugs
a)SCPI modulator attenuation type display instruction error.
b)Switching LF on the LF source page will not refresh the LF icon at the top.
c)TXT file playback error.
d)When playing txt and wdbin files, a local folder is generated in the root directory.
1. Improve the UI.
2. Fix the bug in the trend chart.
3. Modify the number of file systems from 2 to 3 and copy some
configuration files to new location .This will reduce the chance of
a system crash.
4. Need to use transition software to update from old to new
version.
1. Fixed some unit faults in 06version
2. Fixed some bugs about communication with easysdm.
1. Add the customer sensor support.
2. Add the manual calibration support.
3. Optimize flash R&W design。
4. Fix some UI & SCPI bugs
1. Add and optimize the manual calibration function
2. Solve the socket communication in multiple cycles of
open/close, may not respond to the problem
3. Fix the problem that the “READ” command does not
match the description in the programming manual
4. Fix the key test screen no prompt
5. Optimize the connectivity test buzzer response speed
issues
1. Allowed one channel to modulate the other
2. Supported Burst in TrueArb mode
3. Supported Burst Counter in Burst mode so the number of
burst trains can be set to > 1 when trigger source =
external or manual
4. Added an option to let the SDG start with activated
outputs
5. Added 50 ohm output impedance text in the UI
6. Able to remember the digit of a parameter that had been
modified
7. Fixed several bugs:
a) DDS interpolation bug
b) Switching on/off output causes random phase
deviation between channels in Track mode
c) Incorrect square/pulse waveform initialization in
some cases
d) Unable to store the level settings for noise
e) Unable to recognize the command MODE
PHASE-LOCKED
f) Freezing issue when being written a 20MB long file
with the EasyWaveX
g) The setting “BANDSET” with Waveform = NOISE
leads to inconsistent behavior on the Noise
Bandwidth of the output signal; The real-time noise
bandwidth on channel 2 is changed if an AM sine
modulation is switched on channel 1
h) Unable to install the IQ option before 30 trials run out
1. Changed the upper limit of carrier frequency (sine) in
Burst mode from 300 MHz to 500 MHz
2. Added some missing SCPI commands
3. Fixed several bugs:
a) Messed up signal after switching from 500 MHz sine
to square in independent mode
b) Glitches when outputting pulses under certain
circumstances in independent mode
1. Optimized TrueArb algorithm and added 3 types of
interpolation: sinc, sinc27 and sinc13
2. Optimized jitter performance of Square and Pulse
3. Optimized algorithm of Noise generating so the bandwidth
setting lower limit of Noise optimized from 80 MHz to 1
mHz
4. [2017/12/22-174663] Changed the clock switch strategy:
When the clock source is set to External and no actual
external clock signal is being received, the clock source
will not switch to Internal automatically but the clock icon
will indicate that external clock is lost
5. [2017/12/22-174663] Changed default state of 10 MHz
Out to “Disable”
6. Added Slave Delay option in Multi-Device Synchronization
function
7. Added indicator for double function of the soft keys
=>
8. Fixed the following bugs:
a) Crashes when downloading 20 Mpts waveform data
from the EasyWave program
b) Inaccurate burst delay
c) In TrueArb mode enabling AM with large amplitude
may trigger OVP
d) Switching the waveform from IQ to sine with large
amplitude may trigger OVP
e) Switching off the output on the SDG6kX only drops
the output level by approx 20 dB but does not
completely switch it off
1. Removed the warning message when parameter is set to
a value beyond the limit
2. Supported to copy/paste/delete folder in Store/Recall file
manager
3. Optimized initialization time when the waveform is
switched between IQ and other waveforms
4. Supported phase compensation in IQ mode
5. Fixed several bugs:
a) [2018/09/14-924144, 2018/09/14-937446] Signals
get interrupted upon timing-related change of the
settings in Independent mode
b) Anomaly on square/pulse waveform with some
setting
c) [2018/11/06-92208] Anomaly on arb waveform when
toggle the “ArbMode” from DDS to TrueArb and
back to DDS in Independent mode
The first formal release
1. Compatible with Rev.F hardware
2. Support trig dual channel separated when use mannul
trigger.
3. Support using another channel as the modulation
source
4. Support setting the output frequency after the
completion of sweep
5. Counter add fast measurement mode
6. Change the square waveform data used in square FM
modulation
Optimized debounce of the front panel buttons.
Optimized calibration and PV process on the production line.
1. Supported system recovery from U-disk.
2. Fixed a bug which may cause the generator report a
false over voltage alarm when start-up.
1. Changed the amplitude strategy of AM modulation: The
carrier stays fixed at the amplitude, regardless of the
modulation setting.
2. Increased the upper limit of Harmonics number from 10
to 16.
3. Supported Multi-Device Synchronization.
4. Supported Up followed by Down Sweep mode in Linear
sweep (no change for Log sweep).
5. Added an icon to indicate clock source is internal or
external.
Internal: External:
External and the clock is lost:
6. Changed the clock switch strategy: When the clock
source is set to External and no actual external clock
signal is being received, the clock source will not switch
to Internal automatically but the clock icon will
indicate that external clock is lost
7. Supported specifying Arb type by remote control when
the modulating waveform is Arb.
8. Optimized the UI of selecting built-in arbs
9. Supported changing the unit of amplitude without any
amplitude change by long pressing the “Amplitude”
softkey.
10. Added “User” option under “PowerOn Setting”.
11. Added “Normal” as a new UI color style.
12. Added memory space information in Store/Recall UI.
13. Optimized response of the universal knob
14. Added indicator for double function of the soft keys
=>
15. Removed the warning message when parameter is set
to a value beyond the limit
16. Fixed several bugs
a) [2017/04/07-92573] Digit changes when setting
delay of Pulse by the knob
b) Amplitude unit “dBm” is not remembered after a
reboot
1. Allow 0 Hz for CH Frequency In MOD Modes PM, AM,
DSB-AM, ASK, PSK
2. Allow MOD frequency > CH Frequency In MOD Modes
PM, DSB AM, ASK
3. Allow DSB-AM, AM MOD Modes In TrueArb Mode
4. Up Followed by Down Sweep Mode is Supported in
Linear Sweep (No Change for Log Sweep)
5. Added ARB waveform name to display, allowing user to
know what Arb waveform is loaded
6. Fixed some bugs
a) Arb Waveforms In DDS Mode Have One (Small)
Glitch Per Cycle
b) Built in Ramp waveform has a flat spot issue at the
rising slope zero crossing right at the SYNC pulse.
c) Power Up Last Does Not Remember
SRate/Frequency Setting
d) MOD Type = DSB AM Has Amplitude Issues
1. Added statistic function in counter mode
2. Supported using method 2 from our “generating
mulitphase waveforms” application note (refer to
http://www.siglentamerica.com/prodcut-pxzl.aspx?id=9
0&tid=16&T=2) to synchronize 2 units of SDG2000X
3. Fixed some bugs
a) A bug in file system could cause the generator
never startup
b) DHCP does not run at power up
c) Channel lock phase can be wrong in modulation
sometimes
d) Error in calculation of frequency deviation in
counter mode
e) Instrument does not allow CH1 and CH2 to both
use the external input for the modulation source
simultaneously in tracking mode
f) Downward compatibility of .xml state file
g) Unable to delete files with uncommon character(s)
in their names
1. Supported to automatically enable the output
after power-on: Utility | Output Setup | Power
on State
2. Compatible with Rev.E hardware
Fixed the abnormal display issue during start-up on
some units with previous releases. It is not
necessary to update to this version if the unit has
no such a problem.
Fixed the incorrect sweep direction bug
It is strongly recommended to update to the latest
release because in some previous release exists
some serious bug which could cause the generator
never startup
1. Supported system recovery from the U-disk
2. Supported Rev.C hardware with Over Voltage
Protection
Fixed several bugs
a) A bug in P12R1 which could cause the
update to fail
b) Some typo in English help system
• Fix TG switch error at small RBW
• Fix Help and Limit file lost when Factory reset
• Fix Auto open delay unit error in VNA mode
• Fix CNR calc error when no peak found
• Fix OBW result overflow
• Fix Harmonic calc error at 100Hz RBW
• Fix Reflection Open+Short calibration action
• Fix spectrum amplitude error in DMA mode
• Optimize EMI .csv file content to all signal list
• Optimize EMI peak search
• Optimize Auto cal process
• Add Peak search to Spectrum in DMA
Fixed several bugs
a) A bug in file system could cause the
generator never startup
b) Can not wake up from screen saver from
time to time
This version con not be rolled back
• Fix TG switch error at small RBW
• Fix Help and Limit file lost when Factory reset
• Fix CNR calc error when no peak found
• Fix OBW result overflow
• Fix Harmonic calc error at 100Hz RBW
• Fix Reflection Open+Short calibration action
• Fix spectrum amplitude error in DMA mode
• Optimize EMI .csv file content to all signal list
• Optimize EMI peak search
• Optimize Auto cal process
• Add Peak search to Spectrum in DMA mode
New Functionality:
• Add EMI filter in SA for EMI option
Solved Issues:
• Optimize the TG flatness and control procedure
• Optimize Help
• Fix Correction, .csv file, add peak numbers to 20 in EMI mode
• Fix amplitude error in DMA Mode
• Fix upgrade failure issue
New Functionality:
• Add EMI filter in SA for EMI option
Solved Issues:
• Optimize the TG flatness and control procedure
• Optimize Help
• Fix Correction, .csv file, add peak numbers to 20 in EMI mode
• Fix amplitude error in DMA Mode
• Fix upgrade failure issue
New Functionality:
• Update EMI option to a new EMI measurement mode
Improvements:
• Improve TG output harmonic performance
• Update Help file
Solved Issues:
• Fix ADC overload warning
• Fix Ch Power calculation bug in small span
Improvements:
• Add Limit mask offset in SA mode
• TG Level adjustable in Reflection Meas
Solved Issues:
• Fixed AM and FM symbol rate error, Lpf coefficient error, in AMA mode
• Fixed Filter bandwidth error in DMA mode
• Fixed Freq counter error in SA mode
• Fixed Harmonic error in SA mode
• Help Display error
Improvements:
• Optimize trace, marker, signal track, TG normalization,Reflection measurement in SA Mode
• Optimize web server UI
Solved Issues:
• Fixed video trigger fail, errors in Harmonic measurement in SA Mode
• Fixed spectrum amplitude error, meas error, waveform discontinue in MA Mode
• Fixed English help, system message
Notice:
This firmware must be upgraded from v2.1.1.1 or later. If your analyzer has an earlier version, don’t use this FW to update.
Improvements:
Optimize the Ref level and Att couple
Solved Issues:
Fix the zero sweep type when fft in SA mode
Fix the OPC response in EMI mode
1. This firmware must be upgraded from V1.2.8.1 or later. If your SSA3000X
has an earlier version, please update to V1.2.8.1 first and then proceed to
update to the latest version.
2. Fixed bugs on peak search and next peak.
3. Optimize SCPI commands for peak table, screenshot, etc.
4. Optimize the LOG X-scale for EMI test.
5. Optimize the Quasi-Peak detector accuracy for EMI test
1.Long press the left button: one beep sounds,
and the prompt “CC mode prompt tone
disabled” plays.
Long press the left button again: two beeps
sound, and the prompt “CC mode prompt tone
enabled” plays.
1.This firmware is only compatible with
hardware version V6.1.
2.The MCU has been changed to GD32F427.
Firmware versions 1.01.01.02.XXX cannot be
upgraded.
1.Long press the left button: one beep sounds,
and the prompt “CC mode prompt tone
disabled” plays.
Long press the left button again: two beeps
sound, and the prompt “CC mode prompt tone
enabled” plays.
1. Fix: The UI was not updated in time when the power meter was pulled out.
2. Fix: GPIB Address can not be changed.
3. Fix: The mouse is hidden under VNC.
4. Fix: Some wrong SCPI commands were executed.
5. Optimize function with OCXO module.
6. Optimize power meter function.
1. Fix: In the custom mode, when save the state, some items are not saved.
2. Fix: Arb mode and pulse mode conflict. Now they can work simultaneously.
3. Fix: In Arb mode, there are some spurs, when On/Off the RF Mode button.
4. Fix: Some problems about LAN setting, VNC
1. Reduce the overshoot in the pulse function
2. Add pause function and the amplitude can be changed during pause status in
frequency step sweep mode.
3. Added some setting’s SCPI commands.
1. Added user manual calibration and verification function
(option).
2. Added the function that the number of vertical grid can be
set.
3. Added the function of supporting logarithmic display of
vertical coordinates.
4. Added point limit function
1. Added Pulse modulation function.
1. Fixed the range limits for frequency and power in mixer
measurements.
2. Fixed the trace title bar color to match the memory trace
color.
1.Add manual configuration of DC under TDR.
2.Add calibration overwrite function.
3.Optimize synchronization of external triggers for pulse
measurements.
1. 재료 측정 기능 옵션 추가.
2. 스위치 매트릭스 확장 다중 포트 애플리케이션 옵션 추가.
1. Add gain compression measurement and vector mixing
measurement function.
1. Add shortcut buttons at the toolbar.
2. Fix the stuttering problem when adjusting the brightness of
the grid in the waveform area.
3. Solve the problem of probabilistic crashes when opening
dialogs.
4. File storage shortcut keys support more types.
5. Support more power meters.
6. Add picture editing function to the screenshot dialog.
7. Adds peak reference function for bandwidth search.
8. Modify the contents of the Chinese and English help files
9. Modify the user receiver calibration process.
10. The input box will pop up automatically when using the
toolbar button to add Marker.
기능 및 개선 사항:
1. 자동 고정 장치 제거 기능 추가.
2. Added Noise Figure Measurement.
3. Added Power Meter Viewer Dialog.
4. Added support for cal kits with Sliding Load and Offset
Load.
5. Added support for HiSLIP communication protocol.
버그 수정:
1. IF 대역폭 변경 시 발생하는 문제를 수정했습니다.
2. Fixed issue where changing directory in the File Browser
would cause GUI crash.
3. Fixed issue where multi-channel calibration would cause
crash.
4. cal 파일이 제대로 저장되거나 불러오지 않는 문제를 수정했습니다.
5. Fixed issue where sweep time would not estimate correctly
after adding a SA channel
6. Fixed issue with calibration type list display.
7. Fixed a bug where loading S2P file for 2-port de-embedding
would cause crash
8.Fixed issue with calibrating with a DATA-BASED
calibration kit.
9. CSV 또는 SNP 데이터 파일 불러오기 관련 문제를 수정했습니다.
1.Solved the menu button is not updated when the electronic
calibration module is inserted or removed.
2. Fixed a bug that the touch screen may crash under some
cases.
3. Fixed an option installation bug.
1. Added spectrum analyzer option.
2. Fixed input bug in network configuration window.
3. Fixed mouse click input box crash bug.
4. Increased the touch-sensitive area of the DHCP checkbox.
5. Modified the policy when the DHCP request fails.
1. Added independent configuration function of emission
source in spectrum mode.
2. Optimized the sweep process.
3. Optimized TDR interface.
4. Added command for transferring file data.
5. Added marker->SA commands.
1. Added Frequency Offset function.
2. Optimized some functions of spectrum analysis.
3. Added status bar to display smooth status.
4. Added advanced mode for bandwidth search, support
frequency percentage bandwidth setting function.
5. Optimized the Marker->Delay algorithm.
기능 및 개선 사항:
1. 자동 고정 장치 제거 기능 추가.
2. Added Noise Figure Measurement.
3. Added Power Meter Viewer Dialog.
4. Added support for cal kits with Sliding Load and Offset
Load.
5. Added support for HiSLIP communication protocol.
버그 수정:
1. IF 대역폭 변경 시 발생하는 문제를 수정했습니다.
2. Fixed issue where changing directory in the File Browser
would cause GUI crash.
3. Fixed issue where multi-channel calibration would cause
crash.
4. cal 파일이 제대로 저장되거나 불러오지 않는 문제를 수정했습니다.
5. Fixed issue where sweep time would not estimate correctly
after adding a SA channel
6. Fixed issue with calibration type list display.
7. Fixed a bug where loading S2P file for 2-port de-embedding
would cause crash
8.Fixed issue with calibrating with a DATA-BASED
calibration kit.
9. CSV 또는 SNP 데이터 파일 불러오기 관련 문제를 수정했습니다.
1. This firmware must be upgraded from V1.2.8.1 or later. If your SSA3000X has an
earlier version, please update to V1.2.8.1 first and then proceed to update to the
latest version.
2. Improved the stability of sweep and avg detect.
3. Fixed some bugs on sweep, frequncy counter and Log-Log scale.
4. Fixed some bugs on SCPI and the AMK option.
1. This firmware must be upgraded from V1.2.8.1 or later. If your SSA3000X has an
earlier version, please update to V1.2.8.1 first and then proceed to update to the
latest version.
2. Fixed v1.2.8.5 bugs on Refl Mode.
3. Fixed v1.2.8.5 bugs on language and file
1. Fixed several bugs:
a) Add edge trigger for any edge trigger type
b) Bandwidth Limitation in Roll Mode
1. Support adjustment of cursor result display position.
2. Support displaying the average sampling count in the lower right corner of the screen.
3. Support knob acceleration function for cursor value and derivative step.
4. Support setting Math vertical offset and derivative step through the keypad.
5. Optimize trace colors after screen invert (C1 as yellow, C2 as green).
6. Fixed several bugs:
a) Issue where channel invert setting is not saved after reboot..
b) Issue with universal knob indicator light status error.
c) Issue when the menu is hidden, opening History or Search does not show the menu.
d) Issue when measure gate at the screen boundary, the device cannot be powered on after reboot.
e) Issue of missing prompt when adjusting parameters with knob in FFT split-screen mode.
f) Issue where reference label was displayed incorrectly when obscured by menu.
g) Issue of inaccurate AC measurement when inputting large offset AC signal.
Bug Fixes:
1. Fixed issue with Reverse Response Thru Calibration.
2. Fixed issue with log dialog.
3. Fixed issue with recalling state file by SCPI.
4. Fixed issue with abnormal power of SA function under specific conditions.
해결된 문제:
새 기능:
Solved lssues:
해결된 문제:
1.Modified the return of * OPT and added custom IDN and OPT
2.Added LXI
3.Add SCPI command recording function
4.Fixed some bugs
a) Resolve the issue of not reducing the option numbers of uses
b) Resolve the issue of incorrect power in certain situations
1.Display: Supported 100ms, 200ms, 500ms persistence;Supported large font;Optimized the display of custom probe attenuation in the channel description box
2.Fixed several bugs:
a)In Normal trigger mode after one minute the waveform will be refreshed automatically
b)In edge trigger mode Alternating can only trigger on rising edge or falling edge
c)In roll mode level of pattern trigger cannot be set
1. Resolved the issue where SDG800 bandwidth upgrades did not take effect immediately and required a restart.
Fixed some bugs:
1. SCPI instruction configuration exception when inactive channel is set as trigger source.
2. Under the interpolation time base, CSV files need to save the original sampling points, not the interpolated points.
3. After recalling the settings file, sometimes the cursor settings are incorrect.
4. In tracking mode, using the SCPI command to change the source of an X cursor will cause the sources of X1 and X2 to change simultaneously.
5. After setting the reference level for FFT, power down and restart to restore the reference level error.
6. In the average sampling mode, when the time base is modified under a large time base, the waveform does not clear the old sampling points.
7. SCPI instruction reads FFT cursor, returns incorrect unit.
8. Sometimes UART decoding stop bit judgment error.
9. When the unit is mHz, the frequency value read by SCPI is in MHz units, which confuses uppercase and lowercase.
1. Supports webserver (0x6 version hardware); adds virtual panel in web control mode.
2. Screenshots support PNG file format.
3. Adds low-resolution output mode to DC output.
4. Sweep add step mode.
1. Add automatic power-on function.
2. Issues Fixed
1. Add the *RST command..
1. Add the *RST command..
1. Measure: added new items UpLower, △TimeMax, △TimeMin
2. Math:FFT – added windows Blackman-Harris and Gaussian
3. Decode: Added SpaceWire, CANXL
4. Roll: Supported Math in Stop mode
5. Channel: in Eres mode to display the corresponding bandwidth
information in the descriptor box
6. Digital
a) Digital channels ASCII decoding requested
b) Optimized labels’ display
7. Save/Recall:
a) Supported AutoSave
b) File Manager: Supported to Select All
8. WebServer: supported inverted screenshot (according to the Image
style setting in the Save menu); added virtual front panel; added
Run/Stop button
9. UI:
a) Supported custom colors for XY trace
b) Supported to disable Time/Date display
c) Supported to specify Zoom region by drag a rectangle
10. Utility:
a) Supported password for screen saver
b) Key test: Supported to test the rotation direction of the encoders
11. Remote control:
a) a) added commands for counter
b) added commands for △time measurement
12. Fixed several bugs:
a) When saving inverted screenshot as JPG, PNG, trace color of
channels changes
b) SCPI commands don’t work in some cases setting horizontal
parameters
c) Phase measure doesn’t work in some case
d) In Eres mode, abnormal display on trace with 10Mpts length
13. Some SCPI issue on digital channels
Fixed some issue on the production line. From user’s perspective no difference with 1.2.3.1
Optimized Self-cal time from 52 minutes to 22 minutes
1. Fixed several bugs:
a) Add edge trigger for any edge trigger type
b) Bandwidth Limitation in Roll Mode
Made 옵션 16LA 기준
1. Fixed several bugs
a) Cannot exit from Bode plot in some case
b) Cannot detect active probes correctly when the probes are connected at power-on
c) “Set to 50% level” takes the scope frozen in some case
d) Sample rate is displayed incorrectly after reboot from Fixed Memory Length mode
1. Modify the LIST configuration to match EasyPowerX.
Bug Fixes:
1. Fixed issue with Reverse Response Thru Calibration.
2. Fixed issue with log dialog.
3. Fixed issue with recalling state file by SCPI.
참고: 이 펌웨어는 V3.2.2.4.0 이상에서 업그레이드해야 합니다. 분석기가 더 낮은 버전인 경우, 먼저 V3.2.2.4.0으로 업데이트한 후 최신 버전으로 업그레이드하시기 바랍니다.
개선 사항:
해결된 문제:
참고: 이 펌웨어는 V3.2.2.4.0 이상에서 업그레이드해야 합니다. 분석기가 더 낮은 버전인 경우, 먼저 V3.2.2.4.0으로 업데이트한 후 최신 버전으로 업그레이드하시기 바랍니다.
개선 사항:
해결된 문제:
참고: 이 펌웨어는 V3.2.2.4.0 이상에서 업그레이드해야 합니다. 분석기가 더 낮은 버전인 경우, 먼저 V3.2.2.4.0으로 업데이트한 후 최신 버전으로 업그레이드하시기 바랍니다.
개선 사항:
해결된 문제:
1. Decode: Added Spacewire, USB2.0, CAN XL
2. Measure:
a) Added item: UpperLower
b) Added Source: ZFx, ZMx
c) Supported zoom in the measure histogram
3. Cursors: Supported to hide Y cursors in track mode
4. Math:
a) Supported up to 8 traces
b) Added operators: Envelope, tan, atan, FFT Phase
c) FFT:peak added delta display; marker added ordering by number, amplitude and frequency
d) FFT:Added windows Blackma_Harris and Gaussian
e) Supported to hide traces
5. Analysis:
a) SignalScan added decode search
b) Eye Diagram: added Vpp measurement
6. REF: Supported to hide traces
7. Zoom:
a) ZFx supported to set horizontal scale independly; ZFx, ZMx supported to synchronize the display region with Zx
b) Supported to set Zx’ vertical scale with unified adjustment
8. Roll:
a) Supported decode in Stop mode
b) Supported Roll + Zoom in fast display mode
9. Prode: Supported CP150, HVD3106A; Supported to set external scaling of SAP2500D
10. Save/Recall:
a) Autosave supported to save pictures
b) Supported to save all channels’ data to a single file
c) FileManager:Supported to view a picture in full screen mode
11. Display:
a) Axis label supported to set the vertical label on center and right
b) Supported to change the font size
12. System: Time/date supported NTP; Supported education mode
13. Webserver: supported inverted screenshot (according to the Image style setting in the Save menu) ; Added Run/Stop virtual button
14. SCPI: Added command MANual to support to set Amplitude Strategy of Measure
15. Fixed several bugs
a) Sending SCPI commands continuously lead the instrument to frozen state