New Functionality:
Add user calibration for reference clock Improvements:
SA mode: optimize refresh efficiency
AMA mode: optimize AF smoothing effect
NF mode: exporting table results as .csv files
Add FTP switch function Solved Issues:
SA mode: align estimated scan time and regular scan time
BT mode: solve issues including Preset, EDR signal demodulation, and no response to quick operations
PN mode: solve issues including signal averaging failure and incorrect sweep width
NF mode: solve issues including calculation errors
RTSA mode: solve issues including no waveform in screenshots, peak failure,failure to save and load files, and missing reference offset
Add and modify SCPI commands:
1. Period cursor reading error (:CALCulate:MARKer:FCOunt:X?)
2. Query cursor function returns off (CALCulate:MARKer:FUNCtion?)
3. Setting external trigger command is inconsistent with programming manual(:TRIGger[:SEQuence]:SOURce)
4. Harmonic analysis command is missing to obtain each order harmonic frequency and power and obtain total harmonic distortion(:READ:HARMonics:AMPLitude:ALL? :READ:HARMonics:DISTortion?)
5. SA command reads back cent, low, high.Q results when obtaining NDB results(:CALCulate:MARKer:BANDwidth:RESult?)