Fixed some bugs:
1. SCPI instruction configuration exception when inactive channel is set as trigger source.
2. Under the interpolation time base, CSV files need to save the original sampling points, not the interpolated points.
3. After recalling the settings file, sometimes the cursor settings are incorrect.
4. In tracking mode, using the SCPI command to change the source of an X cursor will cause the sources of X1 and X2 to change simultaneously.
5. After setting the reference level for FFT, power down and restart to restore the reference level error.
6. In the average sampling mode, when the time base is modified under a large time base, the waveform does not clear the old sampling points.
7. SCPI instruction reads FFT cursor, returns incorrect unit.
8. Sometimes UART decoding stop bit judgment error.
9. When the unit is mHz, the frequency value read by SCPI is in MHz units, which confuses uppercase and lowercase.
8.1.16 6.1.33/6.1.37R10 Yes
8.1.11 6.1.33/6.1.37R10 No
8.1.9 6.1.33/6.1.37R10 No
8.1.8 6.1.33/6.1.37R10 No
Update instructions
Important!
Because 6.1.33 fixed a bug with self- calibration for each channel. Please perform a self-calibration if the machine is updated to 6.1.33 (or higher)
Very important!
Optimized communication between SDS1xx4X-E and SLA1016. If the SDS1xx4X-E will be upgraded to this version (or higher), the SLA1016 MUST be upgraded to 8.1.16 (or higher) first.
Very important!
Because 6.1.20R1 fixed a bug with self- calibration for each channel, immediately perform a self-calibration if the machine is updated to 6.1.20R1 (or higher)
Fixed the problem: The skew of the two ADCs is not accuracy for hardware of 09 version (System Status shows Hardware Version: 09-xx). The other hardware version has no this problem.