3.1 Check that the trigger source channel is consistent with the waveform being viewed3.2 Manually adjust the LEVEL trigger knob to see whether the waveform is stable.3.3 Use AUTO to adjust the waveform automatically to determine whether the waveform is stable3.4 Adjust time base, rule out false-wave interference (such as alias displayed waveforms)
First, check the power cord and power strip for damage. Check the power switch on the instrument to make sure it is open (SDS1000CFL series has a power switch and a power protection switch). Is the power light illuminated?Is the fan operating? If the fan works, the machine is powered on. If everything is normal…
The term “div” for most SIGLENT specifications is equal to the Vertical Scale value (Commonly shown in units of Volts/div).Take the specification for the DC Gain Accuracy of a SIGLENT SDS1052DL+For ranges ≤ 100 mV/div, the DC Gain Accuracy is given as± [3% × (|reading|+|offset|) +1% × |offset| +0.2 * div+2 mV]If:Reading = 0.5 VOffset is 0.1 VVertical scale…
Information regarding product compliance is available on the datasheet for each product.Contact your local Siglent office for more details if you have trouble finding what you need.You can also write sales@siglent.com
Many SIGLENT instruments feature options that can be added by using an activation code.To generate the activation code:– Purchase the option from your favorite SIGLENT office or distributor. A full list of available options are located at the bottom of the specific product web page.– You will receive an Option Card PDF document from SIGLENT…
The maximum safe voltage of ext trigger channel is 40 Vpp.The specification of working voltages are:Ext : ± 600 mVppExt/5: ±3 Vpp.